Diagnostic Accuracy of Cytochemistry Taking Flow Cytometry as Gold Standard in Diagnosing Acute Leukemia.
DOI:
https://doi.org/10.59058/jaimc.2024.22.1.217Keywords:
Cytochemistry, Sudan black B , Periodic Acid Schiff. stainsAbstract
Background & Objective: Acute leukemia are malignant neoplastic diseases that arise when the
hematopoietic stemcells have malignant transformation into undifferentiated cells permanently. Flow
cytometry is the gold standard to diagnose acute leukemia but it is available at few laboratories among private
and public setups in Punjab and that too is very costly. Cytochemistry is an alternative and relatively cheap
option which is available at most of the setups. So this study was planned with an aim to evaluate the
diagnostic accuracy of cytochemistry using Sudan black B and Periodic Acid Schiff. stains, for acute
leukemia diagnosis in a resource-constrained laboratories.
Methods: After meeting the inclusion criteria and taking appropriate ethical review board approval ,80
patient were enrolled in this cross sectional analytical study. Detailed history and physical examination with
complete blood cells count and numerous hematological indices were obtained by using semi- automated
analyzer; SYSMEX KX-21. Slides were made by using peripheral blood / bone marrow smears, fixed by
methanol and stained for different cytochemical stains. Leukemia cell analysis was done by using the
conventional immunofluorescence method by using the monoclonal antibodies directed against cMPO,
CD2, CD5 , CD7 , CD 10 , CD11 c , CD 14 , CD 19 ,CD 20, CD33 , CD 34 , CD 45 , CD 64 , CD 79a, HLA-DR
, KAPPA, LAMBDA and TdT. The results of flow cytometry and cytochemistry were noted in a pre-designed
proforma and results were entered in SPSS for analysis.
Results: The mean age of patients included in the study was calculated to be 39.40±14.76 years, and male to
female ratio was 1.3:1. The sensitivity, specificity, positive predictive value (PPV), negative predictive value
(NPV) and diagnostic accuracy of cytochemistry using SBB stain for diagnosis of acute leukemia was
91.94%, 100%, 100%, 78.26% & 93.75% respectively taking Flow cytometry as gold standard. Similarly, the
sensitivity, specificity, PPV , NPV and diagnostic accuracy of cytochemistry using PAS stain for diagnosis of acute
leukemia was 100%, 91.94%, 78.26%, 100% & 93.75% respectively taking flow cytometry as gold standard
Conclusion: According to this study, the cytochemistry using both SBB and PAF stains showed good
accuracy for acute leukemia. Therefore, it may substitute Flow ctometry in many cases, in restricted resource
settings.